change "sicky pad" to "sticky pad"
Comment on this formulation: Typical examples for medical devices where the associated functionality of one and the same foot pedal can be changed at runtime - maybe with restrictions - are: The term "can be changed at runtime" can, in my view, be misunderstood, because the assignment cannot be changed during activation. In this respect, I propose to delete "at runtime" in the text.
This could be incorrectly associated with the actual direction of the sky. Alternative: 12 o'clock, 3 o'clock, 6 o'clock, 9 o'clock
text for Device Concepts updated and note added according to todays webex meeting: Device Concepts HF SURGICAL EQUIPMENT is MEDICAL ELECTRICAL EQUIPMENT which delivers HIGH FREQUENCY currents to perform surgical modification of tissue. The most common forms of tissue modification are CUTTING and COAGULATION but may also include tissue ablation, lesioning, shrinkage, sealing or fusion. In today's common HF SURGICAL applications, the HF current is conducted to and from the PATIENT in the following...
P11073-10721™/D1 - Introduction
My proposal using the appropriate wording of -2-2: HF SURGICAL MODE According the defined terms HF SURGICAL MODE is any of a number of OPERATOR selectable HF output characteristics...., such as CUTTING, COAGULATION and the like*
My proposal using the appropriate wording of -2-2: HF application method