Common name: NEUTRAL ELECTRODE's need for PATIENT's skin contact
Description:
Parameter that describes whether the NEUTRAL ELECTRODE needs direct contact to the PATIENT's skin.
Note - Typical "sticky pad" NEUTRAL ELECTRODEs need direct skin contact, while capacitive large-area NEUTRAL ELECTRODEs, which are placed at OR table, do not necessarily need direct skin contact.
Value Set:
change "sicky pad" to "sticky pad"
Diff:
Potentially, there is another way to model this property by using the MODE OF OPERATION pattern:
Common name: NEUTRAL ELECTRODE with need for PATIENT's skin contact mode
Description:
Parameter that specifies that the NEUTRAL ELECTRODE operates in a mode that PATIENT's skin contact is not necessarily needed.
Value Set: MDC_MODE_ON; MDC_MODE_OFF (MDC_MODE_PSD does not make sense in my opinion)
After discussion with @b-andersen on a similar topic: This does not fit to MODE OF OPERATION as defined in IEEE P11073-10701. Thus, I would not prefer this alternative.
It is part of the document as discussed.