# QA script for verification of 3D dose calculation of Oncentra Masterplan
#############################################################################
# Hannes Rennau hannes.rennau@uni-rostock.de

# script reads RTPLAN and RTDOSE file from Oncentra Masterplan and recalculates dose
# based on catheter points and corresponding radiation times for Elekta Ir-192 source
#
# comparisons between dose of commercial treatment planning system (here: Oncentra Masterplan)
# and recalculated dose (as calculated below) are made at identical geometrical
# postions, hence the histogram analysis shown in the output is even more stringent than
# if any gamma-criteria would have been used
#
# in the output simple statistical data is given, showing also radiation time and catheter
# number for verification