QA script for verification of 3D dose of brachytherapy plans
Brought to you by:
rennau
-script recalculates 3D dose distribution of brachytherapy plans from Oncentra Masterplan using RTDOSE and RTPLAN file
The code does not check the current source activity! It recalculates 3D dose based on the RTPLAN data regarding positions and corresponding radiation times and does a scaling of the python-based 3D dose calculation to fit the Oncentra Masterplan dose. Hence this tool does a verification of the 3D dose calculation only.