Hannes Rennau hannes.rennau@uni-rostock.de
QA script for verifiction of 3D dose calculation of Oncentra Masterplan
script reads RTPLAN and RTDOSE file from Oncentra Masterplan and recalculates dose
based on catheter points and corresponding radiation times
comparisons between dose of commercial treatment planning system (here: Oncentra Masterplan)
and recalculated dose (as calculated below) are made at identical geometrical
postions, hence the histogram analysis shown in the output is even more stringent than
if any gamma-criteria would have been used
in the output simple statistical data is given, showing also radiation time and catheter
number for verification