SafetyClassification responsibility
2026-05-11: In chapter 5.3.14, we should add normative content reflecting the opposite direction of responsibility.
Unit of Measure for Ranges in a LimitAlertCondition
2026-05-11: Agreed. Unit of measures should be added to limit alert condition.
Unit of Measure for Ranges in a LimitAlertCondition
2026-05-11: Agreed. Unit of measures should be added to limit alert condition. Make this property mandatory even if the source is a metric wih given unit of measure.
How to handle MaxTechnical/PhysiologicalParallelAlarms
2026-05-11: Can potentially be incorporated into revision. As this appears to be another UI hint, it should not be necessary to have it in BICEPS, though.