We have exactly the same problem: GE monitor stops sending data. The problem can be fixed only by disconnecting and returning the USB cable. We use a USB-COM converter cable on PL2303 chips. Now we are trying to make our own cable that emulates the PL2303 chip, but has the function of automatically disconnecting/connecting the USB interface. When the data from the monitor stops coming, our program sends a special command to such a cable. And he disconnects the USB by himself and connects it again....
Hi Pavan, Have a look at this thread: https://sourceforge.net/p/vscapture/discussion/general/thread/5c39a04b08/ MR400 uses a protocol different from Intellivue. If the waveform protocol is available from the manufacturer, the support can be added. Regards,
Dear John, We have a Philips PMU (MR 400) that has both RS232 port and an ethernet jack. Would I be able to use the VSCapture for collecting the waveforms. At this point the serial data transfer through putty is at 1hz. But we want to actually collect the waveforms and to do so, would you suggest what hardware and software I would need for windows based OS. Thank you for your help. Pavan.
Hi Juan, The Servo-U ventilator numerics and waveforms are fully supported throught the updated VSCaptureCLI app (available in the home page or files section of this site). Details about the complex protocol may be available online by searching the term 'Servo Communication Interface' (SCI) protocol. Regards, John.
Hi Nuri, If VSCapture was useful in your commercial product, please consider acknowledging the project in your site and supporting the project through the donation link. Your solution seems quite interesting. Regards, John.
Hello everyone, I hope you are doing well. I am currently working on a research/engineering project that requires reading real-time data from a Maquet Servo-U ventilator. I have successfully tested the available VSCaptureMaqVent binaries, but I would like to understand in more detail how the communication interface with the Servo-U is implemented. I would like to kindly ask whether it is possible to obtain access to the source code (or partial technical documentation) of the Maquet Servo-U / Servo-i...
Hello John, Our website is www.hl7ebox.com We haven’t added our videos and interfaces yet. However, apart from what you have shared, there is no other information available on the internet about this topic. I will share details about my own project with you via email.
Hello I am developing embedded system too and commercial solution does not help me. I am interested the root cause of this problem. If there is any kinf of teknik details which you can share me, thats will be very usefull for me. thanks
Thank you. Does your embedded system use new drivers for this? Are there any demo videos or sites about it?
Hello Fatih, This issue is quite frustrating. Without unplugging and replugging the USB connection, you cannot re-establish communication — USB converters are a serious problem. I managed to overcome this issue in my own commercial (embedded system) project. I can collect data continuously for 6, 7, even 10 days without interruption. This is a sufficient duration. I receive the packets and send them directly to the server as HL7 messages. hl7ebox If you are considering a embedded commercial solution,...
Hello John, The applications you have published are truly good and useful, and your willingness to help in every matter is something I genuinely appreciate. I would like to express that we were inspired by your work while developing our own embedded system. Wishing you all the best.
Hello Ali, This issue is quite frustrating. Without unplugging and replugging the USB connection, you cannot re-establish communication — USB converters are a serious problem. I managed to overcome this issue in my own commercial (embedded system) project. I can collect data continuously for 6, 7, even 10 days without interruption. This is a sufficient duration. I receive the packets and send them directly to the server as HL7 messages. hl7ebox If you are considering a embedded commercial solution,...
Hello Ali, Have you found a solution to the problem you're experiencing?
Hello John, I'm experiencing a similar problem. Do you have a specific USB to RS232 converter that you use for this? Could please share it?
Yes, my own app. You can adjust that inside the Mindray menu if you want waveforms transfered. I am now working on a history record so still busy...
I have only enabled Mindray HL7 numeric data export, just curious how you got the ECG waveforms, is this your own app or VSCaptureCLI?
Hi Nicolas, From the output (in CSV & JSON url export mode) , it seems the app is not able to connect to the monitor port. Ensure the monitor is set to the correct IP address (check service menu) on the same network & subnet or if the port is not blocked for any other reason. Have you tried running the app with just JSON file export and then uploading the file in another process with tail -f? Alternatively, you can try MQTT export but test the local file export first. Please email xeonfusion at yahoo...
<meta http-equiv="content-type" content="text/html; charset=utf-8">Correct Mindray HL7Sent from my iPhoneOn 25 Nov 2025, at 05:16, John George <xeonfusion@users.sourceforge.net> wrote:Hi Kai, Are the waveforms from Mindray HL7? I plan to add Mindray waveform support in a future version. Regards, John. Support for Mindray Monitors Sent from sourceforge.net because you indicated interest in https://sourceforge.net/p/vscapture/discussion/general/ To unsubscribe from further messages, please visit https://sourceforge.net/auth/subscriptions/...
Thanks you for your wonderful job! I'm using it with a Dräger Vista 120S. I'm afraid that I have an issue with it, when I launch the app like this : /usr/bin/dotnet /var/www/myproject/VSCaptureCLI/VSCaptureCLI.dll --devices 1 --device1type 1 --device1model 4 --device1arg -interval 5 -export 2 -devid Moniteur1 -url https://MYDOMAIN/api/moniteur/probe/1 -port 192.168.70.1 It uses instantly 100% of my CPU and create several processes (even if the client is not connected) (please see attached image)....
Thanks you for your wonderful job! I'm using it with a Dräger Vista 120S. I'm afraid that I have an issue with it, when I launch the app like this : /usr/bin/dotnet /var/www/myproject/VSCaptureCLI/VSCaptureCLI.dll --devices 1 --device1type 1 --device1model 4 --device1arg -interval 5 -export 2 -devid Moniteur1 -url https://MYDOMAIN/api/moniteur/probe/1 -port 192.168.70.1 It uses instantly 100% of my CPU (even if the client is not connected) (please see attached image). Here is the stdout of the CLI:...
Hi Kai, Are the waveforms from Mindray HL7? I plan to add Mindray waveform support in a future version. Regards, John.
Hello Everyone, here is a bit of an update from my side: https://youtu.be/hTx_L5HXtPA next stop is documentation and saving a history.
VSCaptureCLIv1.007 update
Hi Mihir, The most recent data capture app that supports several devices is the VSCaptureCLI available here: https://sourceforge.net/projects/vscapture/files/VSCaptureCLI/ You might be using an older app. For devices using Draeger Medibus.X protocol, select the appropriate module in that app. You should use waveform data capture to collect data at high resolution, such as capturing the Flow waveform. Here's an example of a study using the Draeger MedibusX module https://pmc.ncbi.nlm.nih.gov/articles/PMC9852126/...
Thanks for reporting this, an update fixing this for new Carescape monitors will be released soon.
Hi, firstly, thank you for this application - its great to see! I'm looking to download real time waveform data from a Draeger Evita V500/V800 ventilator at a resolution of atleast 100Hz. I was looking at the application that you've created and v1.003 only collects data at 5s at the most frequent. Is there a way that I can work around that to collect real time data? I need it to be fairly high resolution because I want to track when the patient is inhaling and when they're exhaling for the clinical...
Hi Cooper, Can you send me the screenshots of the EEG waveform from the monitor, the range and the .raw debug file generated by VSCaptureCLI? Please send it to xeonfusion at yahoo dot com. Regards, John.
Hi all we are using VSCaptureWave to capture EEG data from a GE CARESCAPE B450. The exported EEG values seem to be scaled by a factor of 10 compared with what the monitor displays (e.g., the max range should be 500 µV but the max output in the csv file is 5'000 µV). Has anyone encountered this and found a fix or configuration change? Thanks for any pointers! Cooper
An update to this issue. Newer GE Carescape & Canvas monitors may support only higher baud rates, like 115200 and a new DRI 2019 level, unlike legacy GE monitors. An update with this option is planned. The newer monitors also seem to have PL2303 chipset based drivers inbuilt.
An update to this issue. Newer GE Carescape & Canvas monitors may support only higher baud rates, like 115200 and a new DRI 2019 level, unlike legacy GE monitors. An update with this option is planned. The newer monitors also seem to have PL2303 chipset based drivers inbuilt.
Hi, The module for Intellivue monitors is VSCaptureMP in the latest package VSCaptureCLI. Are you using this? The program outputs 2 timestamps: the monitor reported System time and the local time based on the recording PC time. Its better to use the monitor reported time rather than PC time for studies (due to processing lag). Please send me the background of your study for further troubleshooting to xeonfusion at yahoo dot com. Regards, John.
I've noticed that there is a linearly increasing time drift when getting IBP measurements from Intellivue Monitors; if I input raw data into the monitor and then align the output obtained through VSCapture, the drift becomes more and more prominent over time. Why does this occur? Does VSCaptureWave use the PC's internal clock to provide timestamps, or is it a software issue?
S-link is a bad choice :) . I have 50 units. Currently, there is no converter that can provide a stable connection. S-link It shuts down by itself within 2 to 7 hours and will not turn back on unless the USB is unplugged and reinserted. FTDI chipset not supported old version monitors.
I have sent you the documents i have and no there is no service settings related to HL7 unfortunately.
Hi Ali, Based on that HL7 output, it should be possible to communicate directly with the Mindray device via HL7 (as in other Mindray monitors). Can you check the system service settings HL7 tab and share the screenshot? Also PM me any protocol documentation you have at xeonfusion at yahoo dot com. Regards, John.
Hello All, Has anyone tried and retrieved data from Mindray SynoVent E3 Ventilator where there is no gateway or central monitoring system involved. There is a single device which i can make a connection through 4601 port and it sends back MSH|^~\&|||||||ORU^R01|103|P|2.3.1| PID|||ffffffff-ffff-ffff-ffffffffffffffff|| ^ |||M| PV1||I|^^Surgery&1&3232237034&4601&1|||| |||||||||||A| OBR||||Mindray Monitor|||0| OBX||NM|52^||174.0||||||F OBX||NM|51^||70.0||||||F OBX||ST|2301^|| ||||||F OBX||CE|2302^blood...
Hi John, Thanks for your reply! I completely missed that. It's a struggle to find the UC232A with the VID0x0557 but I'm looking for it now! I appreciate your help!
Hi Omar, As mentioned in this thread, a specific older ATEN chipset VID 0x0557 is needed for GE Carescape B650 USB RS232 data export. https://sourceforge.net/p/vscapture/discussion/general/thread/4c2c225ff5/ Other cables may work, but you will need to try one that has its USB driver loaded in the Carescape B450. The monitor is connected via this USB serial to a null modem and in turn to another USB serial cable connected to the laptop. Also make sure that the data export format in Service menu is...
So I connected the B450 to my PC using a serial to USB cable, as required, yet I'm unable to receive any transmissions. I checked on PUTTY, and it seems that both devices are unable to communicate; I'm not getting a response from the monitor. Any help, please?
Hi John, Is there a VSCapture version that reads Mindray SV600 (serial or Ethernet) waveform data?
Hi Ali, This probably has to do with faulty USB serial cables. Try the FTDI based chipsets like Startech null modem cable, that can be used with without a null modem.
Hi Kai, The latest VSCaptureCLI v1.006 version supports Mindray HL7 data capture via LAN. The monitor firmware needs to support HL7 output (as shown in the settings or service menu). The data port configuration can be set to port 4601. It has been tested with uMec series but should also supports other models and OSes like Mac, Windows, Linux. Regards, John
Hello everyone, I’m excited to finally join the Mindray community here. I recently acquired a BeneVision N1 unit from the US, since—as some of you may know—the European firmware unfortunately blocks direct HL7 access (as demonstrated in this video: YouTube link). My goal is to integrate the N1 into a workflow where I can capture and visualize all the data displayed on the monitor—ideally streaming it in real time to another screen or via Wi-Fi to my MacBook. Ultimately, I would like to connect this...
Hello All, We are having a problem keeping data flow up from GE B650 monitors to our Raspberry Pi devices. Some of them take days to encounter a problem, and some of them break communication on the same day. The only solution when/if that happens is to unplug the USB from the monitor side and then plug it back in. Restarting the app or even rebooting the Raspberry Pi device won't help recover the bad connection. Our app that runs on the Raspberry Pi is designed to run multiple devices (ventilators...
Hello everyone, I would love to join the Mindray community here. I just got myself a Mindray N1 from the US as the european firmware blocks access to the HL7 as you can see here https://youtu.be/NBe7d19WzyQ there are ways to get data. I have not yet tried out your software as I am currently only having access to a MAC but hopefully will be on a Windows computer by next week. If anyone has gotten some expereince that might be helpful I would be very glad to learn more. Best from Germany Kai
I agree. But today we have artificial intelligence that can help us do the routine work of reverse engineering. I plan to try to solve this problem.
Hi Romas, Can you private message and email me at xeonfusion at yahoo dot com? Will help you trouble shoot the setup. The VSCaptureGEVent module is part of VSCaptureCLI, so the latter should work with the ventilator. Did you enter 2 devices option, then select ventilator as GE Engstrom and monitor as GE S/5? If you select 1 device in VSCaptureCLI does it output data? Regards, John.
Hello, I understand, that there is no way of using ethernet cable with vscapture solution. But I would like to continue discussion regarding GE ventilator. I would be very kind if someone can share experience of getting data from this device. I did some efforts and would like to ask some questions: 1. I am able to capture some data with VSCaptureGEVent.exe but not with VSCaptureCLI.exe. I need to send json to server, so I need to use software like VSCaptureCLI.exe. I have tried all options when choosing...
Custom folder export maybe offered only to commercial users. Please PM me at xeonfusion at yahoo dot com with your project info and requirements.
Hi, anyone knows how to change the folder where VSCapture stores the generated files? Thank you
Hi Eugeny, Unless we have the X3 protocol doc, it will be difficult to figure it out even if we use apps like Wireshark. Regards,
Hi, John Additional information - our Philps PC runs Philips Hemodynamic Application release 1.3 software. The monitor sends data via direct Ethernet cable to this PC. Do you have any idea, how to recognize and capture which data goes via UDP on this cable ? Best regards, Eugeny
Hi, I want to try to collect data next week from the IP5 (connected to MR400). I also see in the official manual that only 1Hz is possible (via serial and via HIS). Can you tell me how you extracted the data? did you use a standard cable? did you use this VitalSignsCapture software to extract the data? Were you able to capture all data (ECG, PPG, respiratory waveform), or only heart rate for example? Thank you so much for your documents that helps a lot!
Hi Julian, Currently only the S/5 protocol is supported via RS232 or USB port (with connection of ATEN USB serial cable and null modem). The Unity network via LAN is not supported.
Here is a printscreen:
Hi all, I came accross a B40 without the S/5 license. I changed VSCapture to print the serial and did get something, but made no sense: 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...
Hi! The Philips X3 patient monitor is connected via a direct Ethernet cable (point-to-point) to a separate network card of a personal computer running Philips software. I want to receive data from the monitor in parallel. I found this document with protocol description: https://www.documents.philips.com/doclib/enc/fetch/2000/4504/577242/577243/577247/582636/582882/X2%2C_MP%2C_MX_&FM_Series_Rel._L.0_Data_Export_Interface_Program._Guide_4535_645_88011(ENG).pdf As I understand it, the VitalSignsCaptureMP...
Hi! The Philips X3 patient monitor is connected via a direct Ethernet cable (point-to-point) to a separate network card of a personal computer running Philips software. I want to receive data from the monitor in parallel. I found this document with protocol description: https://www.documents.philips.com/doclib/enc/fetch/2000/4504/577242/577243/577247/582636/582882/X2%2C_MP%2C_MX_&FM_Series_Rel._L.0_Data_Export_Interface_Program._Guide_4535_645_88011(ENG).pdf As I understand it, the VitalSignsCaptureMP...
Hi Eugeny, VSCapture has not been tested with X3 separately, it might work once docked with the main monitor. Can you share how you connected it and the protocol document?
Hi I am trying to use VitalSignsCaptureMP to receive data from a Philips IntelliVue X3 patient monitor via LAN. So far it has not worked. The monitor rejects the program's requests on the 24105 portal. Using WireShark, I recorded the data exchange between this patient monitor and the Philips regular computer that receives the data. In the protocol dump I saw that the messages start with the bytes "10 00" and not "E1 00" as expected by VitalSignsCaptureMP. In addition, the data goes between ports...
Thank you for help! I will try that and let you know if I still have issues.
Hi Inbar, The cable and port used is incorrect. You will need to use the DB9 RS232 serial port in the ventilator (image attached) and a usb serial cable. Choose PB ventilator module in the latest VSCaptureCLI app. Please note that the app currently supports numerics only, but waveforms can be added if you can obtain the manufacturer protocol document for this. Please PM at xeonfusion at yahoo dot com. Regards, John.
Hi, I used an ethernet cable connected to the port in the back of the ventilator (image attached). To the ethernet cable I connected the medicollector adapter (https://www.medicollector.com/store/p4/cbl-5010-for-connecting-philips-intellivue-mp20-mp50-mp70-mp90.html), then an adapter from RS232 to USB and that was connected to my PC/Raspberry Pi. This cable configuration worked for me with the Philips Intellivue monitor.
Hi, Can you share your cable configuration and ports used on the ventilator? The latest CLI app contains all the modules you need, theres no need to use other files (which are older versions) Regards,
Hi, I tried to get waveform data from a PB980 vent using this version - VSCaptureCLIv1.004Binary.zip. It did not work, and then I saw you also have this file - VSCapturePBv1.001Binary.zip. I'm not sure how should I work with these files. Should I run only the 2nd one? or the CLI and then the PB specific one? Thank you!
Dear users, If anyone has access to Maquet Servo-i or Servo-U ventilators, please email xeonfusion at yahoo dot com to test a new version of the Maquet module that can capture ventilator waveforms. Thanks, John.
請問一下B105 26PIN 針腳定義
請洽B105 26PIN 針腳定義
As long term user of the project, please consider supporting it through the main page menu link. The GE R860 ventilator is supported by the Engstrom module of VSCaptureCLI under ventilators menu. The issue could be due to the cable wiring. The RS232 ports labelled 5 and 6 on the ventilator are the data output ports. The cable pins to be connected are 2,3 and 5 (see image). Please PM me at xeonfusion at yahoo dot com for further troubleshooting. Regards, John.
Hi John, We are trying to communicate with GE Carescape R860 ventilator with Vscapture software named " GE Engstrom Ohmeda Anaesthesia" under the Ventilators branch. And we are using 3 pin Rx,Tx,Gnd cable for connection. But we cant read anything so far from the Ventilator. Obviously we are doing something wrong, any clues or any help will be highly appreciated. thank you.
The monitor model mentioned is currently not supported.
Hi, John, Does VSCapture work with Dixtal DX2010 monitors?
That's the wrong module being used, it should select Draeger Medibus and not GE. Try downloading the latest version of VSCaptureCLI v1.004 from this site and send me the output. PM xeonfusion at yahoo dot com to help you trouble shoot.
Good day. The capture does not occur. Is it normal that it is saved to the GEVentExportData.csv file, although I select Dreger MedibuGood day. The capture does not occur. Is it normal that it is saved to the GEVentExportData.csv file, although I select Draeger Medibus? The output is belows? The output is below. Tnx. Command line used for GE Engstrom Ohmeda: -port /dev/ttyUSB0 -interval 1 -export 1 -waveset 1 Command line used for CLI: --devices 1 -device1type 4 --device1model 2 --device1arg -port...
Hi Alex, Draeger Atlan data export via the RS232 port maybe supported as it uses the Medibus.X protocol. Choose this option or module in the latest VSCaptureCLI app. Regarding the checksum errors, this usually occurs due to performance issues of the serial cable. Try using a FTDI chipset based USB serial cable for higher data fidelity. BW, John.
Good day, John George Thank you for your work. Please tell me if the data output from the Draeger Atlan is supported? Sometimes when outputting data from some models of ventilators, a message "checksum error". What does this mean? Best regards, Alex Bodrov
Dear John Thank you for the speedy ressponse. I wasn't aware of the new version until yesterday. wecwill most certainly give that a go. We are using the MIB port but will confirm the rest of the details via email many thanks for your help best wishes katie
Dear Katie, Please use the latest version of VSCaptureCLI available from this site, that has the updated data capture modules for several medical devices. Using the LAN port along with a hospital network may lead to data dropouts. Either grant exclusive LAN access or capture data via RS232/MIB port. The LAN port is desirable for data fidelity. Using the option All waves in in the Intellivue module is not recommended, as it can lead to data loss (due to bandwidth constraints of the monitor). It's...
Dear All, We have been using VS capture successfully to collect EEG and physiological data from Phillips Intellivue MX550 monitors via the LAN port & a hospital network. We are using the older MP vs capture version and option 11 (All). Last week it just randomly stopped working. It seems to be collecting pulse ox data fine but not EEG data from the bis module. it is also no longer generating any extra text files? There are no error messages on screen and it appears to be working as one would expect...
Hi Damien, Draeger Vista 120S monitors via LAN are supported, but it uses the "Draeger HL7" export menu option in the VSCaptureCLI app. The Medibus X protocol is used mainly in Draeger ventilators. In Vista 120S LAN network settings, assign a static or dynamic IP address, and use a crossover LAN cable or a straight cable with a router. Vista 120S works once the network is set correctly, also set output format in monitor as HL7. Regards, John.
Hi Damien, Draeger Vista 120S monitors via LAN are supported, but it uses the "Draeger HL7" export menu option in the VSCaptureCLI app. The Medibus X protocol is used mainly Draeger ventilators. In Vista 120S LAN network settings, assign a static or dynamic IP address, and use a crossover LAN cable or a straight cable with a router. Vista 120S works once the network is set correctly, also set output format in monitor as HL7. Regards, John.
Hi, I've test your soft today with a Drager Vista 120S with a RS232 cable plug in the Medibus / X port and my computer on windows 11. When i start a capture with "Command line used for CLI: --devices -device1type 2 --device1model 1 --device1arg -port COM8 -interval 5 -export 1 -waveset 4" a new RAW file is created but no CSV file. I'm attaching the RAW file, is the software compatible with Drager Vista 120S ? Have you a solution ? :D Thank you for you help, Best regards, Damien
Hi, I've test your soft today with a Drager Vista 120S with a RS232 cable plug in the Medibus / X port and my computer on windows 11. When i start a capture with "Command line used for CLI: --devices -device1type 2 --device1model 1 --device1arg -port COM8 -interval 5 -export 1 -waveset 4" a new RAW file is created but no CSV file. I'm attaching the RAW file, is the software compatible with Drager Vista 120S ? Have you a solution ? :D Thank you for you help, Best regards, Damien
Hi Edison, Please PM me at xeonfusion at yahoo dot com for troubleshooting your connection. I think your colleague has already done so? The app definitely works with both devices.
Dear Jhon, we are trying to use your device to connect to an infusion pump and Intellivue Philips monitor MX-80, sadly we have not been succesfull to adquire any signal, we are trying to connect to the monitor through the RS232 port since the LAN port is being used by the hospital. We have made our costum cable according to the Manual as well as changing the configuration setting in the monitor (DCC mode), and we still do not have connection it is for a university project and he have been struggling...
Hello John, I'm a student working in a project that aims at extracting and displaying real-time data simultaneously from 2 devices ( philips intellivue monitor MX850 and fresenius agilia infusion pump). Before going to the ICU in the hospital, I would like to test VScapture simulating the data coming from the monitor. I have a serial port emulator, but I don't have the raw files to simulate the signal ; also I don't know how to use these files with the serial emulator. Do you have some .raw files...
Dear Julian, Could you please share your configuration? I have unsuccessfully tried various pinouts and port settings. Meanwhile, using original B40 multi i/o adapter, everything works fine. Kind regards Romas
Hi John, I am from India, trying to explore a requirement for reading patient monitors, will let you know once the ask takes some shape. Thanks Sathish On Tue, Oct 22, 2024 at 5:19 PM John George xeonfusion@users.sourceforge.net wrote: Please private message your use case and company information, for a professional implementation of the integration solution. VitalSigns Capture CLI source code available? https://sourceforge.net/p/vscapture/discussion/general/thread/418972fce2/?limit=25#3607 Sent from...
Please private message your use case and company information, for a professional implementation of the integration solution.
In all Mindray monitors, please ensure that you enter a MRN before trying to export the data. The monitor may refuse data requests unless a MRN number is filled. Mindray HL7 data export via LAN has been tested succesfully and verified with the VSCaptureCLI app and Mindray uMEC12 monitor.
Use the LAN1 IP settings and manually set the IP address after connecting a crossover LAN cable to your laptop ethernet port. Ensure it is on the same subnet and uses the correct gateway (usually left blank for direct comnection). Alternative is to set IP after sharing the LAN port of your laptop via Wifi. This enables monitor to get the router DHCP assigned IP. Disable Central station (CMS should be in off state) and ADT network setup as in Pg 66 of the operator's manual. https://fcc.report/FCC-ID/ZLZ-WMTSN1/5287411.pdf#page66...
Disable Central station (CMS should be in off state) and ADT network setup as in Pg 66 of the operator's manual. https://fcc.report/FCC-ID/ZLZ-WMTSN1/5287411.pdf#page66 Some of the options are to send unsolicited HL7 data to a server (not the option we want), but what the app actually requests is the solicited HL7 message. For these Mindray models, no additional software may be needed. The app directly requests data using the PDS Gateway protocol.
VS Capture client is a very useful one.. source available for this? also does this support devices other than GE? any documentation available? Very much appreciate this work!