For a precise definition the behavior of devices following the different approaches, we need a definition of the terms in order to use them in requirements.
**First draft: **
ACTIVE OUTPUT TERMINAL-oriented HF SURGICAL EQUIPMENT: The operating concept of the HF SURGICAL EQUIPMENT, including configurable parameters and read-only parameters, is based on the ACTIVE OUTPUT TERMINAL.
Note - This means that for the majority of interactions, the ACTIVE OUTPUT TERMINAL is the leading instance in contrast to the ACTIVE ACCESSORY having a lesser role.
ACTIVE ACCESSORY-oriented HF SURGICAL EQUIPMENT: The operating concept of the HF SURGICAL EQUIPMENT, including configurable parameters and read-only parameters, is based on the ACTIVE ACCESSORY.
Note - This means that for the majority of interactions, the ACTIVE ACCESSORY if the leading instance in contrast to the ACTIVE OUTPUT TERMINAL having a lesser role. The ACTIVE ACCESSORY is not necessarily connected to the HF SURGICAL EQUIPMENT while configuration.
ACTIVE OUTPUT TERMINAL-oriented HF SURGICAL EQUIPMENT: The operating concept of the HF SURGICAL Generator.
The Terminal is direct linked to the generator and independen from the Equipment. This should be stated clear within the definition.
[tcon 05.05.2020]
Add note to definition "ACTIVE OUTPUT TERMINAL-oriented HF SURGICAL EQUIPMENT":
Note 2 – The ACTIVE OUTPUT TERMINAL is always a part of the HF SURGICAL EQUIPMENT.
[tCon 05.05.2020]
Consolidated version of the definition
ACTIVE OUTPUT TERMINAL-oriented HF SURGICAL EQUIPMENT: The operating concept of the HF SURGICAL EQUIPMENT, including configurable parameters and read-only parameters, is based on the ACTIVE OUTPUT TERMINAL and more or less independent from the applied ACTIVE ACCESSORY.
Note 1 - This means that for the majority of interactions, the ACTIVE OUTPUT TERMINAL is the leading instance in contrast to the ACTIVE ACCESSORY having a lesser role.
Note 2 - If there are no ACTIVE ACCESSORY identification functionalities, the ACTIVE OUTPUT TERMINAL-oriented approach is completely independent from the actually attached or intended to be attached ACTIVE ACCESSORY.
If there are ACTIVE ACCESSORY identification functionalities, the ACTIVE OUTPUT TERMINAL-oriented approach can be influenced by the actually attached ACTIVE ACCESSORY. This might lead to restrictions of Output Control Settings.
Note 3 - The ACTIVE OUTPUT TERMINAL is always a part of the HF SURGICAL EQUIPMENT.
[tCon 19.05.2020]
Consensus decision: accept this draft --> doc-me
Text acknowledged by experts.