#37 May be allowing the association between instrument (VITRUAL TERMIAL) and ACTIVE OUTPUT TERMINAL if the device cannot recognize the actual type of the instrument.

5.0: HH 20-04
wont-fix
Martin
2020-04-25
2020-01-21
Martin
No

Some / most device can only be recognized as monopolar / bipolar, however not whether it is a tweezers, scissors etc.

Whenever the device cannot recognize, the OPERATOR has to configure this. -> maybe provide this configuration via SDC.

Discussion

  • Stephan Klöckner

    As long as the instrument provides not the information by itself, we should only limit this to the possible HF-mode. There is no need for the configuration of the instrument type for the standard devices.

     
  • Martin

    Martin - 2020-03-25
    • assigned_to: Martin
    • Milestone: 4.0: TÜ 20-01 --> 5.0: HH 20-04
     
  • Martin

    Martin - 2020-04-25
    • status: open --> wont-fix
     
  • Martin

    Martin - 2020-04-25

    [tCon 07.04.2020]
    The situation is valid for instruments without identification chip. However, when somebody plugs in such a device, the instrument selection will be done directly at the device. We do not see a use case for an SDC-based external control operation.

    -> won't fix

     