#37 May be allowing the association between instrument (VITRUAL TERMIAL) and ACTIVE OUTPUT TERMINAL if the device cannot recognize the actual type of the instrument.
As long as the instrument provides not the information by itself, we should only limit this to the possible HF-mode. There is no need for the configuration of the instrument type for the standard devices.
If you would like to refer to this comment somewhere else in this project, copy and paste the following link:
[tCon 07.04.2020]
The situation is valid for instruments without identification chip. However, when somebody plugs in such a device, the instrument selection will be done directly at the device. We do not see a use case for an SDC-based external control operation.
-> won't fix
If you would like to refer to this comment somewhere else in this project, copy and paste the following link:
As long as the instrument provides not the information by itself, we should only limit this to the possible HF-mode. There is no need for the configuration of the instrument type for the standard devices.
[tCon 07.04.2020]
The situation is valid for instruments without identification chip. However, when somebody plugs in such a device, the instrument selection will be done directly at the device. We do not see a use case for an SDC-based external control operation.
-> won't fix