reported pressure in daily screen and records beyond practical limits
Open Source CPAP Research and Review Software
Status: Beta
Brought to you by:
jedimark
I imported data from my s9 which is bipap, the pressure reported by the software on the daily screen seems like it is a bit too high, higher than what it actually is. Attached is a screenshot.
this is the text shown on the daily screen below the ahi reading:
VPAP ST(S9)AMER
PAP Mode: APAP
Pressure: 340282346638528859811704183484516925440.0-0.0cmH2O
Hi,
I'm fairly certain this issue has already handled in source code, and the fix should be in the next (0.9.7) release.
Although I would like to double check it with some VPAP ST data, as I do not have any on hand to verify this.
Would it be possible to email me a zipped copy of your SD card?
/Mark
Yeah I could send you the data, I am not sure what your email address is though. Also do you have PGP and/or GPG for email encryption?