Hi Juan, VSCaptureCLI latest release should work with Draeger Evita ventilators, provided you set the MedibusX protocol in the settings menu of the ventilator. You may need to set the baud settings in the ventilator as 19200-8-Even-1 and try a straight DB9 cable (or a DB9 null modem cable if that doesn't work). Regards, John.
The Draeger Infinity protocol is supported by VSCaptureCLI Draeger Infinity module. The models like C500 should work with appropriate cables from the manufacturer.
Hi, Waveform export is not enabled via MQTT or JSON HTTP currently in the free public version (if its for commercial use, drop me a mail at xeonfusion at yahoo dot com). A workaround is to use the CSV waveform export and pipe the output. Regards, John.
Hi Emre, you need to order the 14 pin cable (p/n 5206441) from Draeger. The output is through X5 or X3 port of the Infinity docking station (IDS). The wiring diagram is different for X3 and X5. Regards, John.
Great work building this full stack solution!
See: https://sourceforge.net/p/vscapture/blog/2026/06/vscapturecli-v1008/
VSCaptureCLI v1.008
Hi Chris, The SystemLocalTime is the time on the decoding PC/RPi, while the Time and RelativeTime in the Intellivue module are from the monitor device reported time. VSCapture has no limitations in the rate of decoding data, the monitor CPU may get overloaded by requesting data at too frequent intervals like 1 second. Ensure that you set a data query time interval like 10 seconds or more. Please send me your configuration, cable used, data interval etc to xeonfusion at yahoo dot com for further troubleshooting....