Not clear if this clause also applies to Pre-sets and settings. E.g. can the consumer use this information to determine the resolution for a setting from the StepWidth in order to limit user input at the UI to a certain resolution for a potential remote change operation?
TBD
This clause applies to Pre-sets and settings, however it is still questionable if TR0239 and TR0240 shouldn't be expanded to pre-sets and settings as well.
TR0239 and TR0240 do not apply on settings and pre-settings as those can be enumerations as well which come without a technical range. Does it make sense to expand both to settings and pre-settings and at the same time limit the METRIC type to numeric, RTSA and DSA?
Last edit: David 2021-07-13
See discussion in [#25].
Related
Closed: P11073-10701 Comments:
#25Also related: [#23]
Related
Closed: P11073-10701 Comments:
#23Structure improved and introductions added.
@peterkranich: Please check after publication of next draft revision.