#152 Clarify that non-usage of StartTime and StopTime within RealTimeSampleArrayMetricState

Corrigendum
closed
David
2024-05-07
2020-06-25
Martin
No

AbstractMetricValue/@StartTime and @StopTime are useless for RealTimeSampleArrays. Thus, this should be clarified by a note.

Discussion

  • Stefan Schlichting

    • status: unread --> open
    • assigned_to: Stefan Schlichting
     
  • Stefan Schlichting

    I agree for continous waveforms, but not for intermittent measured RTSA esp. if there is a post processing after the data was captured between start and stop time.

    Proposal: Won't fix

     
  • Björn Andersen

    Björn Andersen - 2022-03-25
    • Milestone: Corrigenda --> Revision
     
  • David

    David - 2022-03-28
    • status: open --> accepted
    • assigned_to: Stefan Schlichting --> David
     
  • David

    David - 2022-03-28

    Add a note that for RTSA there is usually no start and stop time, but there may be cases in which a start time is used for a purpose specific to the metric type, e.g. Area Under Curve where start time depicts the start time of calculating the area.

     
  • David

    David - 2023-04-03
    • Milestone: untriaged --> Corrigendum
     
  • Björn Andersen

    Björn Andersen - 2024-03-18
    • labels: --> RealTimeSampleArrayMetricState
     
  • Björn Andersen

    Björn Andersen - 2024-03-18
    • status: accepted --> solved
     
  • Björn Andersen

    Björn Andersen - 2024-03-18

    Agreed on the note that @d-gregorczyk proposed. No change to model or requirements needed as RTSAs may carry a start/stop time, e.g. for ECG snippets.

     
  • David

    David - 2024-04-11
    • status: solved --> in-progress
     
  • David

    David - 2024-04-11
     
  • David

    David - 2024-05-07
    • status: in-progress --> closed
     