I have some trouble with PDF creation. Whenever I use a landscape format e.g. ThreeXFourPlusOne V4 to V6 are blank and the rhythm strip stops at that point too. All waveforms are displayed in the portrait format though
This is a very late response, but somehow I never received any e-mail
notification about my Support Request on sourceforge. I guess you already
forget about this problem, but still would like to handle this support request:
This issue looks to be very intressing. My first guess is that this issue is caused, because you are using an ECG with right instead of the standard left layout. Replacing V3, V4, V5, V6 with V3R, V4R, V5R and V6R.
I am very intrested in resolving this issue even though it is almost a year later. Could you perhaps send me a copy of the ECG file that had this issue? If so I will provide you with my personal e-mail adres.
If you would like to refer to this comment somewhere else in this project, copy and paste the following link:
Oh no I took a look at the provided pdf. This issue is caused by something else. The file you used is shorter then 10 seconds long. To be precise the file must be 7,5 seconds long (this becomes clear with the lead II at the botoom). V4-V6 are missing because there is no signal available for second 7,5 till 10 seconds of the reconding.
If you would like to refer to this comment somewhere else in this project, copy and paste the following link:
This is a very late response, but somehow I never received any e-mail
notification about my Support Request on sourceforge. I guess you already
forget about this problem, but still would like to handle this support request:
This issue looks to be very intressing. My first guess is that this issue is caused, because you are using an ECG with right instead of the standard left layout. Replacing V3, V4, V5, V6 with V3R, V4R, V5R and V6R.
I am very intrested in resolving this issue even though it is almost a year later. Could you perhaps send me a copy of the ECG file that had this issue? If so I will provide you with my personal e-mail adres.
Oh no I took a look at the provided pdf. This issue is caused by something else. The file you used is shorter then 10 seconds long. To be precise the file must be 7,5 seconds long (this becomes clear with the lead II at the botoom). V4-V6 are missing because there is no signal available for second 7,5 till 10 seconds of the reconding.