ECG Viewer - Gain setting
ECG Toolkit support for: SCP-ECG, DICOM, HL7 aECG, ISHNE & MUSE-XML
Brought to you by:
mvanettinger
is it just me or does gain setting in ECGView not work correctly.
at 5mm/mv one square should be 200uv, or five squares should be 1mv or 1000uv? Right?
Also is calibration pulse correct? Seems like when changing to gain other than 10mm/mv grid and calipers are not correct. Am I wrong?
Is it the gain of data that changes and not grid view? Is this how ECG machines do it?
Thanks
Two gains same grid measurements
I guess you are right about this bug in the caliper.
Must be a very small fix so probably can deliver a patch tomorrow.
I solved the bug you described. You can download the patch here:
http://sourceforge.net/projects/ecgtoolkit-cs/files/ecgtoolkit-cs/ecgtoolkit-cs-2_3/patch-20130404/